Last updated on 19-Jun-2001 12:09 AM

 

 

Cryostat Assembly

1 Before Starting

2 Installing the Cryogenic cabling (300K to 80K)

3  Installing the Support Trusses and Radiation Shield

4   Installing the Main Optical Bench

5   Installing the Closed Cycle Cooler

6   Installing the Joule-Thomson Cooler

7  Installing the Slit/Field Rotator Assembly

8   Installing the Front Optical Bench

9   Installing the 4K Arm

10   Installing the Detector Box and Array

11   Installing the FOB Radiation Shield and Vacuum Cover

12   Installing the Grating Drum and associated vacuum cover

13   Installing the Calibration Unit

14  Telescope Install

 

 

 

1 Before Starting

The Assembly of Michelle requires the lifting and handling of heavy assemblies and usually requires the presence of at least two and preferably 3-4 people. All such staff should be conversant with safe handling procedures and be equipped with safety boots and helmets, at least one should have had appropriate slinging training.  The assembly is based on the use of the Michelle Handling Manipulator and an associated hoist (at UKIRT the jib crane).The whole process is also eased by ready access to storage facilities for the unassembled modules and appropriate (wheeled) methods of moving these modules within range of the lifting facilities.

 

Durations:
A complete assembly from scratch as described below will take several days. As a planning guide I have given an estimated time for each major subsection at the heading level.

 

2.   Installing the Cryogenic Cabling (300K to 80K)

3.   Installing the Support Trusses and Radiation Shield

 

4.   Installing the Main Optical Bench (90 mins)

DATE/TIME:
4.1   Turn the vacuum vessel until the interface with the front optical bench is horizontal and at the top. Lower the main optical bench whilst guiding the nitrogen fill pipes through the holes in the vacuum vessel until it is a few inches above the truss feet.

Figure 4.1   Lowering the main optical bench into the vacuum vessel, with the fill tubes being guided through the radiation shield.

 

Comments:Done by:
Checked by:

4.2  Fit the collet clamps to the nitrogen fill tubes and push them into the holes in the radiation shield.

 

Comments:Done by:
Checked by:

 

4.3  Lower the main optical bench the final inches onto the truss feet and tighten the eight M16 x 70mm bolts to 140Nm.

Figure 4.3 Tightening the bolts holding the MOB onto the truss.

Comments:Done by:
Checked by:
4.4 Tighten the six bolts holding the collet clamp pairs on each of the two nitrogen fill tubes onto the radiation shield.

Figure 4.4 Bolting down the collet clamps from the MOB nitrogen fill tubes. Note that the Radiation Shield Heater panel shown here now has clamping bars across the backing plate.

Comments:Done by:
Checked by:
4.5 Install the vacuum fittings and the supporting collars for the MOB nitrogen fill tubes where they enter the vacuum vessel. Finally add the white plastic protection sleeves.

Figure 4.5   Installing the support collars on the nitrogen fill tubes.

 

Comments:Done by:
Checked by:

 

5   Installing the Closed Cycle Cooler (20 mins)

DATE/TIME:
5.1 Lower the cooler into the vacuum vessel (Figure a) and tighten the six mounting bolts (Figure b).

This is a solid copper ring now.

Figure 5.1a. Lowering the CTI1020 cold head in the vacuum vessel. (Note that the copper wick assembly has now been replaced by a solid copper sleeve)

 

Figure 5.1b. Bolting the CTI1020 cold head onto the vacuum vessel.

 

Comments:Done by:
Checked by:

 

5.2  Tighten the eight screws which mount the the copper collar on the radiation shield to the copper sleeve attached to the 1st stage of the coldhead. These should be torqued to M6 setting.

(NEW PHOTO REQD)

Figure 5.2.  The second stage adapter plate can be seen at the end of the 2nd stage column, protruding through the radiation shield. The wicking bundles go from the adapter plate to the copper block which is bolted to the main optical bench at the bottom of the picture. The circular collar at the top clamps the radiation shield onto the sleeve from the first stage. The first and second stage temperature sensors are shown.

Comments:Done by:
Checked by:
Fit the superinsulation collar around the copper sleeve from the first stage of the coldhead and secure with lacing tape.

 

5.3  Install the Radiation Shield temperature sensor on the copper plate behind the field rotator motor and attach connector.  (This may have not been removed as now clear of area.)

Comments:Done by:
Checked by:

 

5.4  Tighten the six 6-32UNC screws and their invar spacers which hold the 2nd stage adapter plate onto the end of the 2nd stage pillar (to required torque setting). This is an awkward operation requiring a short Allen key.

Comments:Done by:
Checked by:

 

5.5  Install the temperature sensor on the second stage adapter plate and attach connector.

Comments:Done by:
Checked by:

 

5.6  Install the 2nd stage wicking between the adapter plate and the main optical bench (see Figure 7). The screws are M5 with a torque setting of ...

Comments:Done by:
Checked by:

 

6.   Installing the Joule-Thomson Cooler (90 mins)

DATE/TIME:
6.1  Rotate the cryostat to a position where the Joule-Thomson can be wheeled into place horizontally on its trolley. If fitted, remove the blanking plates in the centre section of the vacuum vessel and radiation shield, and carefully guide the J-T cooler into the cryostat.

Figure 6.1.   Installing the Joule-Thomson cooler in the centre section of the vacuum vessel. Note that it now comes fitted with it's own Radiation Shield which will obscure the inner pipework. This operation will be done in Hawaii using the Handling Trolley.

 

Comments:Done by:
Checked by:
6..2  Install the electrically insulating washers and inserts and then tighten the bolts connecting the cooler to the vacuum vessel.

Figure 6..2. The electrically insulating inserts which screw the Joule-Thomson cooler into the centre section of the vacuum vessel.

 

Comments:Done by:
Checked by:
6.3  Fit the o-ring around the interface at the radiation shield and secure in place with the split ring.   Access to the rear section requies removing the connector plate as shown.

position of the electrically insulating moat.

Figure 6.3 The interface between the J-T cooler first stage and the radiation shield. The green circle marks the position of the electrically insulating moat. (NEW PHOTO req'd)   Check that the J-T cooler is electrically isolated from the cryostat.  Fit the small flat shield to the end of the J-T radiation shield.

Comments:Done by:
Checked by:

 

7   Installing the Slit wheel/Field Rotator Assembly (20 mins)

 

Orientate the cryostat to have the coldheads on the underside per figure.  Remove the perspex cover covering the Collimator entrance, these screws hold in the slit/FRot assembly.  You may need to loosen the screws retaining the transverse light baffles in the Main Optical Bench are loose.  Make sure the status cables for both Field Rotator and in particular the Slit Wheel are attached to the assembly.   Check also that light path through array is clear.
Comments:Done by:
Checked by:
Fit the Assembly into the Main Optical Bench making sure it is pushed snugly into it's "corner" before fitting and securing the three retaining bolts. Note that the proximity of the light baffling at the assembly's motor plate can lead to an improper location.   Take care to avoid damaging cabling on the radiation shield.
Comments:Done by:
Checked by:
If necessary tighten the screws securing the transverse light baffles between the Slit/Field Rotator assembly and the Nitrogen tank. Fit the thermal wicks to each motor, connect the two motor cables at both ends and terminate the status cables at the Radiation Shield end.

Comments:Done by:
Checked by:

Fit tape along gaps at the rearmost of the motor plate as this is a known light leak area.

 

8   Installing the Front Optical Bench

DATE/TIME:
8.1  The filter wheels are light enough to be lowered into the slot in the FOB by hand as shown below. The FOB is supported as described in Section 8.2. The two motor cables, a status cable and the temperature sensor on one Filter wheel should then be reconnected.

Figure 8.1   Lowering the filter wheel assembly into the front optical bench. (Alignment features highlighted)

Comments:Done by:
Checked by:
8.2  Remove the perspex covers over the field rotator optics and the opening to the spectrometer camera through the Nitrogen can.
Comments:Done by:
Checked by:
8.3  Referring to Figure 12, rotate the instrument such that the mating face of the MOB is parallel to the FOB as it hangs on the single support. Using the trolley facility on the hoist move the FOB into engagement with the MOB - the FOB feet help with the alignment but manual guidance is also required. Attach the FOB to the MOB by partially tightening the six bolts, three per side.

Figure 8.3. Mounting the Front Optical Bench onto the Main Optical Bench.

Comments:Done by:
Checked by:
8.4  Rotate the instrument to an orientation where the weight of the FOB helps pushes it into the corner defined by the three reference pads shown in Figure 13. The six bolts(with Invar washers) will need to be loosened first for positioning before being tightened (M8, 36.2Nm).

Reference padReference padReference pad

Figure 8.4. Aligning the FOB to the MOB. The three reference pads are circled in green in the lower picture; the filter wheel assembly and fore-optics are not shown for clarity.

Comments:Done by:
Checked by:
8.5  Connect the motor thermal wicks at the filter wheel motors. Replace the long side panel covering the joint between FOB and MOB.   This should then be taped up.
Comments:Done by:
Checked by:
8.6  Connect all six stepper motor cables (making sure the control racks are depowered) and one datum status cable that interface the FOB to the cabling on the centre section of the radiation shield.
Comments:Done by:
Checked by:

9.   Installing the 4K Arm

10.   Installing the Detector Box and array

 

DATE/TIME:

The detector in its gold coloured box should normally be mounted as late as possible in the assembly procedure to avoid damage. It should be stored in a protected, dry atmosphere and carried in a container that keeps it electrically contacted to ground through an approximately 1 million Ohm resistance (to prevent fast changes in potential difference and hence high instantaneous current flow).

 

10.1  Start off by installing four 1 million Ohm grounding paths running from the detector focus and translation mechanism to locations as listed below, and as shown in Figure 10.1.
  1. To the top of the Joule-Thomson cooler, which should in turn be connected to the telescope ground.
  2. To a conducting mat placed on which the person installing the detector box should stand.
  3. To a wrist strap on the installer.
  4. To the grounding stud on the gold detector box.

    Figure 10.1. The detector box and translation mechanism grounding scheme for electrically safe installation.

Comments:Done by:
Checked by:
10.2  Mount the detector box into the translation mechanism using the four screws, ensuring that the two are electrically isolated from each other using the G10 tube inserts, and drawing film under the copper washers (as shown in Figure 10.2).

Figure 10.2. Detail of the mounting screw arrangement for the detector box.

Comments:Done by:
Checked by:
10.3  Use a voltmeter to check that the detector box is isolated from the optics.
Comments:Done by:
Checked by:
10.4  Attach the cover to the detector box, feeding through and screwing down the four D-type connectors.
Comments:Done by:
Checked by:
10.5  Connect the detector cables without delay. If this is not possible, ensure that the detector box's voltage cannot float, for example using the scheme shown in Figure 10.5.

Figure 10.5. Grounding the detector box when the detector cables are not connected.

Comments:Done by:
Checked by:
10.6  Install the silver foil covered radiation shield over the detector housing as shown in Figure 10.6.

Figure 10.6. Installing the Detector Box's 25K Radiation Shield

Comments:Done by:
Checked by:
10.7  Securely bolt the copper arm shown in Figure 18 onto the '4K' stage of the Joule-Thomson cooler. Attach the two sets of copper wicks leading from it, one to the detector cold finger assembly and the other to the detector box.

Figure 10.7. The copper arm from the J-T Cooler to the detector box.

Comments:Done by:
Checked by:
10.8  Connect the cable assembly, running along the copper arm from the J-T stage to the detector box, as shown in Figure 10.8.

Figure 10.8. The detector cable assembly mounted on the copper cooling arm.

Comments:Done by:
Checked by:
10.9 Next, build up the 4K radiation shielding around the 4K stage of the J-T cooler and the copper arm (Figures a and b). Note that this must not come in thermal contact with any elements which are being cooled to 25K.

Figure 10.9a and 10.9b. The double layer aluminium 4K radiation shield foil radiation.

Comments:Done by:
Checked by:
10.10 The final appearance of the detector box and J-T cooler is shown in Figure 10.10.

Figure 10.10. The assembled detector box, with the 4K cooling circuit installed.

Comments:Done by:
Checked by:

 

11 Installing the FOB Radiation Shield and Vacuum Cover (UKIRT) (45 mins)

DATE/TIME:
11.1  Before closing the cryostat front end the system should be connected to the Cryostat Control system to confirm the performance of the various Front end mechanisms. A further check is required to confirm that the Temperature sensors are connected. (A complete status switch check can be performed using the Status cable break-out box if desired).

11.2  Mechanism Check
Using CCS dm screens do a force datum on the mechanisms in turn.   One person should observe the mechanism itseld for motion, noise etc., whilst the other should observe the dm screen for datum switch operations etc.
Before driving the motors remove all perspex films used to cover the motor shafts.

Mechanism

Test 1

Test 2

Test 3

Test 4

Slit Wheel
Field Rotator
Filter Wheel A
Filter Wheel B
Imager Inject
Imager Extract
Focus Drive
Translation Drive

Comments:Done by:
Checked by:

11.3  Temperature Sensor Check

Indicate temperature reading either directly from scanners or from CCS

Temperature Monitor A

Temp (K)

Temperature Monitor B

Temp (K)

1 Rad Shield Nitrogen Can   1 JT Stage 1
2 GM Stage 1   2 JT Stage 2
3 Rad Shield Grating Cover   3 JT Stage 3
4 GM Stage 2   Unused
5 Rad Shield Front Unused
6 Grating Assembly   6 Filter Wheel Motor
7 Slit/Field Rotator housing Unused
8 Imager Stop Unused

 

Lakeshore 330

Temp (K)

A Array cold block  
B Array substrate  

Comments:Done by:
Checked by:

11.4  Remove all protective covers and shields from the optical path and then mount the front section of the radiation shield, securing the quick-release latches. Now fit and tighten the 16 screws (with belleville washers) to a torque of 4 Nm, followed by the four screws which secure the copper wick from the central radiation shield (also 4NM torque).
Comments:Done by:
Checked by:
11.5  Attach the temperature sensor to the front radiation shield, accessible through the hole in the Shield and close off the hole.
Comments:Done by:
Checked by:
11.6   Install the front section's MLI blanket.. Check for electrical short circuits between the Joule-Thomson cooler and the cryostat, preferrably for a range of gravitational orientations of the instrument.
Comments:Done by:
Checked by:
11.7   Remove the protective gasket from the vacuum tight surface of the centre section of the cryostat.
Comments:Done by:
Checked by:

11.8   Dismount the front section of the vacuum vessel (mass = 220kg) from its storage plate and lower it onto the centre section. This is best done with the interfacing surface tilted approx 17 deg (top of Radiation Shield horizontal).   Fit the two alignment dowels that fit the front cover (and hence the calibration unit) to the centre section.

Figure11.8. Installing the front cover of the cryostat (UKIRT configuration).

Comments:Done by:
Checked by:
11.9 Tighten the bolts (M12 * 45 long) holding the section of the vacuum vessel onto the centre section.   The cryostat will need to be rotated to access those nearest the manipulator.
Comments:Done by:
Checked by:
Remove the Window Assembly from it's dessicated storage. Remove the external cover to fit the assembly to the Vacuum cover. Then replace the cover to keep the window dry.    This should be done only when the cryostat is ready for pump.
Comments:Done by:
Checked by:

 

12 Installing the Grating Drum and associated Vacuum cover

DATE/TIME:
12.1   Position the cryostat so that the grating drum can be lowered from above at the top with the interface between the grating drum and the centre section horizontal.
Comments:Done by:
Checked by:
12.2
Comments:Done by:
Checked by:
12.3   Remove the protective plastic film covers from the stepper motors.
Comments:Done by:
Checked by:
12.4   If it has been stored with the main drum bearing vertical, the grating drum can be rotated using two hoists and slings as shown in Figure 11.4. The purple coloured sling is use to lower the drum into the cryostat, and it should have two, equal length brothers.

Figure 12.4. Rotating the grating drum.

Comments:Done by:
Checked by:
12.5 If absent, fit five grey coloured bar shaped snubbers to the centre section radiation shield to act as guides when inserting the drum).
Comments:Done by:
Checked by:
12.6 Fit two lifting shackles (the third is not required) to the grating drum on the side nearest to the motors and lift into position over the cryostat opening. First remove the perspex cover from the Grating Assembly and then the perspex cover from the open face of the centre section where it protects the spectrometer collimator and camera optics. Lower the Grating Exchange Assembly into the centre section as shown in Figure 11.6, (drum mass 126kg).    Three spring fingers push the assembly into a three point mount in the last 20mm of wave.

Figure 12.6. Lowering the grating drum onto the main optical bench.

Comments:Done by:
Checked by:
12.7 Tighten the captive bolts holding the grating drum to the centre section to a torque of 67Nm using the special tools provided (two Allen key ended bars 57cm and 67cm long). Remember to replace the access hole covers after use.
Comments:Done by:
Checked by:
12.8 Plug in the two motor cables and one status cable running from the centre section to the grating drum. Note that the tag screw is absent from the cover of the wavelength drive motor since it would foul the radiation shield.
Comments:Done by:
Checked by:
12.9 Connect the two thermal wicks between the motors of the grating drum and the Radiation Shield

The wick attached to the motor which rotates the drum is tricky to access (see Figure 12.9).

Figure 12.9 The thermal wicks on the grating drive motors. The hand is pointing to the wavelength drive wick at the left (note the grey snubber bar on the radiation shield). The right hand picture shows the difficult to access grating exchange motor wick.

Comments:Done by:
Checked by:

Grating Assembly Mechanism Check

Before closing the grating vacuum cover the system should be connected to the Cryostat Control system to confirm the performance of the Grating mechanism and status switches. The connection to the Grating temperature sensor should also be checked. (A complete status switch check can be performed using the Status cable break-out box if desired). Before driving the motors remove all perspex films used to cover the motor shafts.

The test of the mechanisms is probably best performed as a series of "forced datums" selecting each grating in turn. The status switches should be observed throughout to check for operation.

Drive

Echelle

Med N1

Med N2

Low Q

Low N

Round 1          
Round 2          

Grating Assembly Status Switches

Status Switch set

1

2

3

4

Wavelength Datum     NA NA
Grating Datum (spatial datuming)     NA NA
Wavesafe (can wavelength be selected)     NA NA
Drumsafe (can drum be rotated)     NA NA
Gratcode (to identify the grating)        

Comments:Done by:
Checked by:

Grating Assembly Temperature Sensor Check

Indicate temperature reading either directly from scanner or from CCS

Temperature Monitor A

Temp (K)

Temperature Monitor B

Temp (K)

1 Rad Shield Nitrogen Can
NA
1 JT Stage 1
NA
2 GM Stage 1
NA
2 JT Stage 2
NA
3 Rad Shield Grating Cover
NA
3 JT Stage 3
NA
4 GM Stage 2
NA
Unused
5 Rad Shield Front
NA
Unused
6 Grating Assembly
6 Filter Wheel Motor
NA
7 Slit/Field Rotator housing
NA
Unused
8 Imager Stop
NA
Unused

Comments:Done by:
Checked by:

12.10 Remove the five grey snubber bars.
Comments:Done by:
Checked by:
12.11 Manually lower the Grating end Radiation Shield cover into position. Ensure that the small side cover has been fitted (removal no longer required). Tighten the bolts connecting the grating radiation shield to the centre section radiation shield (torque should be 14.7 Nm).
Comments:Done by:
Checked by:
12.12 Install the multi-layer insulating blanket over the grating drum radiation shield, securing the joint with the velcro pads.
Comments:Done by:
Checked by:
12.13 Mount eye-bolts in the grating drum cover and lower it (mass = 100kg), onto the centre section of the vacuum vessel. (Note that it is easy to lower into position 180 deg out)

Figure 12.13. Installing the grating drum vacuum cover.

Comments:Done by:
Checked by:
12.14 Tighten the bolts connecting the grating drum vacuum cover to the centre section.
Comments:Done by:
Checked by:

 

13 Installing the Calibration Unit

DATE/TIME:
13.1 The calibration unit relay optics (the left-most silver box of Figure 13.1) should be bolted on to the cryostat first, and the shutter and slide mechanisms slid over the tube protruding from the relay optics from the right.

Figure 13.1 The Calibration Unit (The Integrating Sphere is not fitted)

Comments:Done by:
Checked by:

14 Installing the Detector Electronics

DATE/TIME:
14.1 The on-cryostat EDICT array controller electronics rack slides onto the guides shown in Figure 14.1 and is held in place by the two hex bolts in the slide end stops.

Figure 14.1 Mounting the On-Cryostat Detector Control Electronics Rack

Comments:Done by:
Checked by: